Agilent 1200 Series Nano HPLC System with G1315B DAD

Agilent 1200 Series Nano HPLC System with G1315B DAD
Item ID:

1200-Nano-G1315B


Condition: Refurbished

Availability: In Stock

Manufacturer: Agilent (HP)Price:

Request Quote/ Offer

Warranty Included, Onsite/ Remote Installation and Training Available

Shipping:

CAN/US: 3-5 business days
International: 5-10 business days



For assistance and further information:

Toll Free: 1-866-777-3688
Tel: 905-946-8066
E-Mail: info@spectralabsci.com

Spectralab is your one stop shop for all your analytical laboratory needs!

Your source for

PRE-OWNED

and Spectralab Scientific

CERTIFIED

Equipment.

We will match our competitor's price!

To Sell, Trade in, or Service Your Equipment

This HP/ Agilent 1200 Series Nano HPLC System comes with the followings:

 

This system is in great condition. It has been tested at SpectraLab Scientific. A 90 day warranty is included.

We encourage you to contact us to speak with one of our expert and professional team members. Thanks for looking and we look forward to hearing from you!

 

Performance Specifications:

G2226A Nano Pump
Type Specification
Hydraulic system Two dual piston in series, with proprietary servo-controlled variable
stroke drive, floating piston, active inlet valve, solvent selection valve
and electronic flow control for flow rates from 0.1 to 1 µl/min
Settable column flow range 0.01 – 4 µl/min
1 – 2500 µl/min (with the electronic flow control bypassed)
Recommended column flow range 0.1 – 1 µl/min
200 – 2500 µl/min (with the electronic flow sensor bypassed)
Optimum composition range 1 to 99% or 5 µl/min per channel (primary flow), whatever is greater
Composition precision < 0.2 % SD, at 500 nl/min (default settings),
Minimum primary flow/pump channel is 5 µl/min
Delay volume Typically 300 nl from the electronic flow control to the pump outlet for flow rates up to 4 µl/min. For flow rates up to 4 µl/min and electronic flow control active: primary flow path 180 – 480 µl; system pressure dependent (default settings; calculated volume)Typically 180 to 480 µl (system pressure dependent) for flow rates up to 2.5 ml/min. (default settings; calculated volume)
Pressure range 20 to 400 bar (5880 psi) system pressure
Compressibility compensation User-selectable, based on mobile phase compressibility
Recommended pH range 1.0 – 8.5, solvents with pH < 2.3 should not contain acids which attack stainless steel. Upper pH range is limited by fused silica capillaries.
Control and data evaluation Agilent ChemStation for LC
Analog output Communications For pressure monitoring, 2 mV/bar, one output
Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready,
start, stop and shut-down signals, LAN optional
Safety and maintenance Extensive diagnostics, error detection and display (through control module and Agilent ChemStation), leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas.
GLP features Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of seal wear and volume of pumped mobile phase with user-settable limits and feedback messages. Electronic records of maintenance and errors.
Housing All materials recyclable.
G1379B Degasser
Type Specification
Maximum flow 5 ml/min per channel
Number of channels 4
Internal volume Typically 1 ml per channel
pH range 1-14
RS232 output For diagnosis purpose
Material in contact with Solvents PTFE, FEP, PEEK
G1377A Micro WPS
Type Specification
GLP features Early maintenance feedback (EMF), electronic records of maintenance and errors
Communications Communications Controller-area network (CAN). RS232C, APG-remote standard, optional four external contact closures and BCD vial number output
Safety features Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display
Injection range 0.01– 8 µl in 0.01 µl increments with the small loop capillary 0.01–40 µl in 0.01 µl increments with the extended loop capillary
Precision Typically < 0.5% RSD of peak areas from 5 – 40 µl, Typically < 1% RSD from 1 – 5 µl Typically < 3% RSD from 0.2 – 1 µl
Pressure range 400 bar (5880 psi)
Sample viscosity range 0.2 – 5 cp
Sample capacity 2 × well-plates (MTP) + 10 × 2 ml vials
108 x 2-mL vials in 2 x 54 vial plate plus 10 additional 2 mL vials
30 x 6-mL vials in 2 x 15 vial plate plus 10 additional 2 mL vials
54 Eppendorf tubes (0.5/1.5/2.0 mL) in 2 x 27 Eppendorf tube plate
Injection cycle time Typically < 30 s using following standard conditions: Default draw
speed: 4 µl/min Default eject speed: 10 µl/min Injection volume:
0.1 µl
Carry-over Typically < 0.05% using the following conditions: Column:
150 x 0.5 mm Hypersil ODS, 3 µm Mobile phase:
Water/Acetonitrile = 85/15 Column Flow rate: 13 µl/min Injection
volume: 1 µl caffeine (=25 ng caffeine), 1 µl water to test carry over
Outside wash of needle before injection: 20 sec with water using flush port
G1316A TCC
Type Specification
Temperature range 10 degrees below ambient to 80 °C
up to 80 °C: flow rates up to 5 ml/min
Temperature stability ± 0.15 °C
Temperature accuracy ± 0.8 °C
± 0.5 °C (with calibration)
Column capacity Three 30 cm
Warm-up/cool-down time 5 minutes from ambient to 40 °C
10 minutes from 40 – 20 °C
Dead volume 3 µl left heat exchanger
6 µl right heat exchanger
Communications Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN via other 1200 series module
Safety and maintenance Extensive diagnostics, error detection and display (through control module and Agilent Chem Station), leak detection, safe leak handling, leak output signal for shut down of pumping system. Low voltages in major maintenance areas.
GLP features Column-identification module for GLP
documentation of column type
Housing All materials recyclable
G1315B Diode Array Detector
Wavelength range 190-950 nm
Light source Deuterium and tungsten lamps
Detection type 1024 – element photo diode array
Short term noise (ASTM) single & mult-wavelength ± 1° x 10-5 AU at 254 and 750 nm
Drift 2° x 10-3 AU / hr at 254 nm
Linear absorbance range > 2 AU (upper limit)
Wavelength accuracy ± 1 nm
Wavelength bunching 1 – 400 nm
Diode width < 1 nm
Flow cells Standard: 13µl volume 10 mm cell path length and 120 bar (1760 psi) pressure maximum
Semi-Micro: 5µl volume, 6 mm cell path length and 120 bar (1760 psi) pressure maximum
Micro: 2µl volume, 3 mm cell path length and 120 bar (1760 psi) pressure maximum
High Pressure: 1.7µl volume, 6 mm cell path length and 400 bar (5880 psi) pressure max.
80 nano: 0.08µl volume, 100 mm cell path length and 50 bar (725 psi) pressure max.
500 nano: 0.5µl volume, 10 mm cell path length and 50 bar (725 psi) pressure max.
Analog outputs Recorder / integrator: 10 mV or 1 V, output range
0.001 – 2 AU, two outputs
Control and data evaluation Agilent Chem station for LC
Slit width 1, 2, 4, 8, 16 nm
Communication types CAN, GBIP, RS-232C, APG, Remote: ready, start, stop and shut down
Wavelength range 190-950 nm
Light source Deuterium and tungsten lamps
Detection type 1024 – element photo diode array
Short term noise (ASTM) single & mult-wavelength ± 1° x 10-5 AU at 254 and 750 nm
Drift 2° x 10-3 AU / hr at 254 nm
Linear absorbance range > 2 AU (upper limit)

 

Request Quote:

*We do not share your information*

Name
Company
Address (City, State)
Country
Email
Email Confirmation
Phone
Your Budget/ Offer
Enquiry

Quality
Our refurbished instruments are tested and repaired to meet factory specifications. We follow Installation Qualification, Operational Qualification, and Performance Qualification protocols (IQ, PQ and OQ).

Expertise
Our team of skilled technicians are experts at fixing analytical instruments. We have been working with Agilent, Waters, Perkin Elmer, Sciex, Shimadzu, and Thermo instruments for many years. We are constantly updating our knowledge with the latest technology and information to enable us to deliver the best support in the industry. We provide value added support and qualification/validation with all our instruments. We endeavor to develop strategic partnerships with our clients so that they consider our team as an extension of their employees. We are an affordable alternative to many normally expensive business projects.

Warranty and Support
We offer warranty on all refurbished instruments. Warranty starts at 3 months, with the option to extend the period. While your instrument is on warranty, we will provide Email (support@spectralabsci.com) and phone support and troubleshooting if you require assistance. If there is ever anything wrong with the equipment, we will replace it. Warranty begins from the date your equipment is successfully delivered. Any issues that occur during warranty should be reported immediately, describing the issues in as much detail as possible. Do not attempt to repair the equipment or involve a 3rd party without our express permission – this will result in voiding our warranty.

We will match our warranty duration with our competitors.

Installation and Training
If you purchase equipment from Spectralab Scientific, we offer installation and familiarization by our service technicians for an additional fee. If the instrument is particularly complicated, we highly recommend this service. If you’ve purchased less-complex equipment from us that you would like to install yourself, we can provide Team Viewer, email, or telephone support for no additional cost.

Large Inventory
We have a large inventory of equipment, components and parts in stock. You can find all the equipment you need for you lab in our store, as well as mix and match components for your systems.

International Shipping
We have sufficient export experience, including professional packing and proper documentation. Complementary shipping and handling can be applied upon immediate purchase. We include customs fees, brokerage, and duties for shipping to the US.

Price Matching & Discounts
Purchasing our refurbished instruments can save you anywhere from 30% to 90% compared to purchasing brand new. We will try to match or beat any price from our competitors, so feel free to inquire! Complementary shipping and handling can be applied upon immediate purchase.

Featured Products

  • New
    Shimadzu LCMS-8060 CL Triple Quadrupole Mass Spectrometer

    Shimadzu LCMS-8060 CL Triple Quadrupole Mass Spectrometer


    Condition:

    Refurbished

    Request QuoteView
  • New
    Agilent 8890 Gas Chromatograph G3540A

    Agilent 8890 Gas Chromatograph G3540A


    Condition:

    Refurbished

    Request QuoteView
  • NEW Thermo Dionex ASE 350 Accelerated Solvent Extractor


    Condition:

    Brand New

    Request QuoteView