The epMotion 5075 LH is the perfect solution for more complex pipetting tasks. It offers the same precision and accuracy as the epMotion 5070, making it the ideal liquid handling robot for demanding applications such as the nucleic acid purification with magnetic beads or routine applications with several plates. The 12 lid positions, option of automatically changing pipetting tools, and the optional plate gripper make the epMotion 5075 LH the most flexible automated liquid handling system.
The epMotion 5075 LH has 12 locations for labware (racks, plates, thermoracks and thermoblocks or tips). Up to three thermal modules can be installed. It comes with an attached Integrated PC and epBlue 20 Version software.
Features
- 12 ANSI/SBS microplate positions
- Maximum pipetting accuracy from 1 to 1,000 µL
- Automatic exchange of all tools (gripper, 1-channel & 8-channel)
- Optical sensor for detecting liquids, labware, tips
- Compatible with tubes (0.5 mL to 50 mL) and microplates with up to 384 wells
- Fully enclosed, compact housing with door lock
- Detection of tip size and tip quantity
Technical Data
Gripper carrying capacity | 1,200 g |
Operation | Electronic |
Pipetting type | Air cushion systems |
SLAS/ANSI positions | 12 |
Thermal module cooling down period of heating/cooling plate | from 25 to 4 °C in 5 minutes |
Thermal module heating off period of heating/cooling plate | from 25 to 95 °C in 8 minutes |
Thermal module setting range | 0-110 °C |
Application | Nucleic acid purification |
Volume range | 0.2 µL – 1 mL |
Power supply | 100 – 240 V ±10 %, 50 – 60 Hz ±5 % |
Max. power consumption | 80 W |
Dimensions (W × D × H) | 107 × 61 × 67 cm / 43 × 24 × 27 in |
Weight w/o accessories | 85 kg / 187.4 lb |
Conductor – X, Y, Z Positioning
Random measurement error | ±0.1 mm |
Systematic measurement error | ±0.3 mm |
Detector
Optical sensor | Contact-free recognition of fill level, inserted tools, labware, tip types and quantities |
Dispensing tool
Random measurement error (1 µL)1,2 | ≤5 % |
Systematic measurement error (1 µL)1,2 | ±20 % |
Random measurement error (50 µL)1,2 | ≤0.4 % |
Systematic measurement error (50 µL)1,2 | ±1.0 % |
Random measurement error (1,000 µL)1,2 | ≤0.15 % |
Systematic measurement error (1,000 µL)1,2 | ±0.7 % |
1in pipetting mode, free jet, without pre-wetting, with distilled water, at 20 °C
2See Eppendorf Application Note 168 for typical pipetting performance