The 7000D Triple Quadrupole GC/MS is the latest version of the most successful GC/MS/MS in history. The 7000D is now even easier to use and more efficient than ever, due to its new Dynamic MRM (dMRM) mode of acquisition. Coupled with the 9000 GC, the 7000D MS works in perfect harmony to enhance productivity, save resources and alert operators about pending maintenance.
System includes:
- G7000D Triple Quadrupole GC/MS/MS (EI)
- Intuvo 9000 GC
- 7693A Liquid autosampler with 150 position Sample tray
- PC with data and Library
- Rotary vaccum pump
Type | Specification | |
---|---|---|
Mode of operation | EI standard, CI optional | |
Ion source material | Noncoated, proprietary inert source | |
Ion source temperature | 150 to 350 °C | |
Filaments | Dual filaments for EI | |
Source cleaning | Automated and vent-free with patented (or proprietary) JetClean option | |
Electron energy | 10 to 300 eV | |
Mass filters (2) | Proprietary monolithic hyperbolic gold-coated quadrupole | |
Mass axis stability | < ± 0.10 u over 24 hours (10 to 40 °C) | |
Quadrupole temperature | 106 to 200 °C | |
Mass range | m/z 10 to 1,050 | |
Resolution | Selectable, 0.7 to 2.5 Daltons, default tune Settable, 0.4 to 4.0 Daltons, custom tune | |
Scan rate | Up to 20,000 u/s | |
Tuning | Autotune or manualDetector Triple-Axis HED-EM with extended-life EM and dynamically ramped iris | |
MRM speed | 800 transitions/sec | |
Minimum MRM dwell | 0.5 msec | |
Collision cell | Linear hexapole | |
Collision cell gas | Nitrogen with helium quench gas Collision energy Selectable up to 60 eV | |
Vacuum system | Dual stage turbomolecular pump Total gas flow up to 8 mL/min | |
Software | Agilent MassHunter acquisition, data handling (quant/qual) and reporting Pesticides and Environmental Pollutants MRM database with over 8,000 optimized transitions (optional) |