Specifications |
|
Optics |
Single beam principle with simultaneous measurement across the photodiode array |
Light source |
Deuterium lamp |
Wavelength range |
200nm – 344nm (UV-range) 345nm – 595nm (VIS-range) |
Pixel bandwidth |
1.9nm (UV-range) 3.3nm (VIS-range) |
Wavelength accuracy |
± 0.75nm (UV-range) ± 1.5nm (VIS-range) |
Noise |
< ± 0.8 x 10-5AU, after 3h warm-up at 254nm, time step 1sec., empty flow cell |
Drift |
< 5 x 10-4AU/h, after 3h warm-up at 254nm, empty flow cell |
Flow cell volume |
Standard flow cell: 10µl, 9mm path Standard flow cell/PEEK capillaries: 10µl, 9mm path Preparative flow cell: 6µl, 2mm path Micro capillary flow cell: 0.14µl, 8mm path |
Flow cell pressure rating |
Standard flow cell: 100bar (10,000kPa) Standard flow cell/PEEK cap. 100bar (10,000kPa) Preparative flow cell: 100bar (10,000kPa) Micro capillary flow cell: 400bar (40,000kPa) |
Measuring range |
0AU to 2AU |
Time constant |
0.1sec |
PC-connection |
Proprietary cable |
Data channels (UVD 170S) |
4 with CHROMELEON™ as of version 4.10 |
This refurbished Dionex UVD 170S UV/VIS Detector is in great condition. It has been tested at SpectraLab Scientific.
A 90 day warranty is included.
We are seasoned professionals in this area and we would be happy to have you speak with one of our experts here at SpectraLab.
Thanks for looking and we look forward to hearing from you!