Type |
Specification |
GLP features |
Early maintenance feedback (EMF), electronic records of maintenance and errors |
Communications |
Communications Controller-area network (CAN). RS232C, APG-remote standard, optional four external contact closures and BCD vial number output |
Safety features |
Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display |
Injection range |
0.1 – 100 µl in 0.1 µl increments Up to 1500 µl with multiple draw (hardware modification required) |
Precision |
Typically < 0.5% RSD from 5 – 100 µl, Typically < 1% RSD from 1 – 5 µl variable volume |
Sample viscosity range |
0.2 – 5 cp |
Sample capacity |
2 × well plates (MTP) + 10 × 2 ml vials 100 x 2-ml in one tray |
Injection cycle time |
Typically < 30 s using following standard conditions: Default draw speed: 200 µl/min Default eject speed: 200 µl/min Injection volume: 5 µl |
Carry-over |
Typically < 0.01% using the following conditions: Column: 125 x 4 mm Hypersil ODS, 5 µm rate: 1 ml/min Injection volume: 1 µl caffeine (1 mg/ml), 5 µl water to test carryover Outside wash of needle before injection: 20 sec with water using flush port |
This refurbished HP / Agilent 1100 Series G1367A AutoSampler is in great condition. It has been tested at SpectraLab Scientific.
A 90 day warranty is included.
We have special expertise on analytical equipment. We encourage you to contact us to speak with one of our expert and professional team members.
Thanks for looking and we look forward to hearing from you!
Multiple available