PerkinElmer TurboMatrix Headspace Sampler 40

PerkinElmer TurboMatrix Headspace Sampler 40
Item ID:

PE-G0-0420


Condition: Refurbished

Availability: In Stock

Manufacturer: PerkinElmerPrice:

Request Quote/ Offer

Warranty Included, Onsite/ Remote Installation and Training Available

Shipping:

CAN/US: 3-5 business days
International: 5-10 business days



For assistance and further information:

Toll Free: 1-866-777-3688
Tel: 905-946-8066
E-Mail: info@spectralabsci.com

Spectralab is your one stop shop for all your analytical laboratory needs!

Your source for

PRE-OWNED

and Spectralab Scientific

CERTIFIED

Equipment.

We will match our competitor's price!

To Sell, Trade in, or Service Your Equipment
Specifications
Hardware Sample capacity: TurboMatrix 40 – Forty vials with intelligent overlapping thermostatting of up to twelve vials.
Sample vials Use PerkinElmer 22 mL sample vials (P/N B0104236) for a maximum liquid sample volume of 15 mL. With the low-volume vial adapters (P/N N6120110) the TurboMatrix 16 and TurboMatrix 40 systems can also use 9 mL sample vials (P/N N9302134) for a maximum sample volume of 2 mL. Note: Use of non-PerkinElmer sample vials and patented cap enclosures is NOT recommended and system performance problems resulting from the use of non-PerkinElmer vials and caps will NOT be covered under the system warranty.
Sample thermostatting Solid aluminum thermostatting oven. Thermostatting period and temperature precisely controlled for each sample. Software algorithm automatically optimizes sample thermostatting overlap.
Sample temperature Temperature settable from 35°C to 210°C in 1°C increments or can be run at ambient.
Sample needling Stainless Steel (Platinum/Iridium needle optional).

Temperature settable from 35°C to 210°C in 1°C increments or can be run at ambient.

Transfer line length Choice of 1008 mm or 1650 mm.
Transfer line Temperature

 

settable from 35°C to 210°C in 1°C increments or can be run temperature at ambient.
Transfer line system Deactivated fused silica capillary transfer line or capillary column (0.32-µm i.d. or 0.25-µm i.d.) shielded inside a heated transfer tube between the headspace autosampler and GC injector.
Column compatibility Compatible with all capillary column diameters in split and splitless (direct coupled) mode. On-column sampling using 0.25-mm i.d. or 0.32-mm i.d. capillary columns and 1/8-inch o.d. packed columns.
Sampling method Pneumatic, pressure-balanced sampling (no syringe or gas sample loop). Sample vials are pressurized with carrier gas. During the injection period, carrier gas flowing to the GC system is replaced by pressurized sample gas. Rapid transfer of analytes is ensured without re-equilibration in a gas sampling valve or syringe and subsequent loss of analytes. Injection amount is programmable by time or by volume without requiring any hardware change.
Analyte compatibility Polar and non-polar organic compounds can be analyzed without any change of internal tubing. Low risk of sample composition change during analyte transfer. Analysis of chemically active compounds such as free volatile organic acids, sulfur compounds, amines and other nitrogen containing organic compounds can be performed using the optional Platinum/Iridium needle.
Pneumatic control

 

Choice of:

— Manual pneumatics using built-in pressure regulator (0-60 psig).

— Electronically controlled Programmable Pneumatic Control (PPC) module (0-60 psig).

Modes of operation: Constant mode

Routine analysis mode with constant, equal thermostatting time for each sample. Includes Express Step Function to automatically optimize oven movement for maximized sample throughput.

MHE mode

Multiple headspace extractions from each vial. Up to nine extraction steps with intermediate vent performed automatically. For use in method development and validation or for routine analysis of special samples.

Progressive mode

Method development tool used in determining required equilibration time and in performing kinetic studies. Thermostatting time is automatically increased to determine optimal thermostatting time.

Control Stand-alone

Each headspace sampler is a self-contained stand-alone unit. All instrument functions are easily operation controlled from the touch screen of the built-in graphical user interface. Set and actual values may be displayed at any time. Configurable user interface for simplest possible interaction.

— Simple one-touch operation

— Single method operation

— Sequence-based multimethod operation with up to eight methods

— Password protection

— Ten operating methods can be stored in the instrument

— Parameters remain stored with last used method directly available after power-up

Operating languages English, French, German, Italian, Japanese, Spanish and Russian.

Software control

An optional control software package is available to control the headspace samplers from a Microsoft Windows® environment. The headspace control software offers graphical representation of instruments and parameters with intuitive interaction. Sequence building follows the easy drag-and-drop principle. A runlog tracks and saves information on samples and performed analyses for GLP compliance. Context sensitive on-line help is available for all functions.

Gas chromatograph/data system interfacing Output signal lines for READY and RUN to start gas chromatograph and data handling devices.

Input lines for START, STOP and READY to control the headspace unit externally. Six built in relays (4 with contact closures, 2 with 24-volt outputs).

Economy mode Programmable power-saving and gas-saving feature with automatic wake-up function.
Priority vial One or two vial positions can be assigned to priority vials. Urgent vials can be placed in these positions and will be inserted into the analysis sequence for quickest possible analysis. Available for the TurboMatrix 40 and TurboMatrix 110 headspace samplers, non-trap and trap models.
Maintenance alarms Settable injection counter that alerts user when it is necessary to perform routine maintenance.
System integrity checks Automatic system leak checking

Static leak check: Checks the whole system for leaks prior to analysis

Dynamic leak check: Checks each vial for leaks as it is sampled

Vial temperature calibration

User calibration supported.

Optional accessories Shaker

Patented frequency scanning shaker for quicker sample equilibration during thermostatting. All vials are agitated simultaneously to allow quicker equilibration. Automatic frequency scanning

program optimizes the agitation process. Available for the TurboMatrix 40 and TurboMatrix 110 headspace samplers, non-trap and trap models.

Cryofocusing/water trap

On-column cryofocusing accessory for preconcentration of analytes and improved detection limits. Water trap included to allow large-volume sampling from aqueous samples without ice blockage. Available for the TurboMatrix 40 and TurboMatrix 110 headspace samplers, non-trap models only.

BCD output

Injector adapter kits

Injector adapter kits are available for most contemporary GCs. Side-mount kit available for PerkinElmer AutoSystem/AutoSystem XL Gas Chromatograph.

Removable sample tray

Extra sample trays for vials to be loaded into while instrument is running. Available for the TurboMatrix 40 and TurboMatrix 110 headspace samplers, non-trap and trap models.

AutoSystem injector interface

Special hourglass injector insert to connect the fused silica transfer line from the headspace sampler to the capillary column provides the performance advantages of a direct connection while still maintaining the convenience of connecting to a split injector.

Physical details Ambient operating conditions

Temperature range: 15 °C to 32 °C

Humidity: 75% maximum RH without condensation

Power requirements: 100/120/240 V; 50/60 Hz; 900 VA

Dimensions in cm TurboMatrix 40 (non-trap and trap): W*D*H 43*61*61
Weight TurboMatrix 40: 33 kg
TurboMatrix Headspace Sampler with Trap Hardware

Inline adsorbent trap Material: Quartz

Length: 132 mm

Outer diameter: 3.5 mm

Inner diameter: 2.8 mm for adsorbent packing with 0.7 mm inlet/outlet channel

Packed bed length: 25 mm maximum

Trap load temperature: 5°C above ambient to 100°C

Desorption temperature: 5°C above ambient to 400°C

Heating rate: 300°C/min or 2400°C/min linear program

Heating hold time: 0 to 999 minutes

Desorption pressure: 0 to 60 psig

Desorption pressure time: 0 to 999 minutes

Desorption split: Off or fixed at 10 mL/min

Cooling rate: 400°C to 40°C in less than 2 minutes

Trap materials:

Standard: Tenax GA or Air Toxics

Custom: To order

Empty: For packing by user

Thermal runaway protection: From firmware and by thermal fuse

Sample pathway Diameter: 0.5 mm

Valving: Pressure balanced (no mechanical valve in sample flow path)

Operation Pressure pulsed Vial pressure:

0 to 60 psig (independent of dry purge and column pressure)

extraction Vial pressurization per cycle: 0.1 to 99.9 minutes

Trap load time per cycle: 0.1 to 99.9 minutes

Trap load flow rate: 50 mL/min fixed

Number of pulse cycles: 1 to 4

Water management Technique: Dry purge

Time: 0.1 to 99 minutes

Temperature: 5°C above ambient to 99°C

Pressure: 0 to 60 psig (independent of vial and column pressure)

Purge flow rate: 50 mL/min fixed

Column isolation Standard on all trap models. Allows the HS system to be serviced while the GC and flow feature detector are still active.

Flow rate: 1 to 50 mL/min by manual mass flow controller.

Internal standard: A gaseous standard mixture is used to charge a gas sampling loop and that is subsequently addition option delivered to the sample vial prior to sampling. May be ordered as a factory option or as a field upgrade to trap-based instruments.

Loop size: 0.5 mL

Loop load flow rate: Adjustable by needle valve

Loop load time: 0 to 2 minutes

Loop pressure equilibration time: 0 to 2 minutes

Loop inject time: 0 to 2 minutes

This refurbished PerkinElmer TurboMatrix Headspace Sampler 40 is in great shape. It has been tested at SpectraLab Scientific. A 90 day warranty is included.

We are seasoned professionals in this area and we would be happy to have you speak with one of our experts here at SpectraLab. Thanks for looking and we look forward to hearing from you!

 

Request Quote:

*We do not share your information*

Name
Company
Address (City, State)
Country
Email
Email Confirmation
Phone
Your Budget/ Offer
Enquiry

Quality
Our refurbished instruments are tested and repaired to meet factory specifications. We follow Installation Qualification, Operational Qualification, and Performance Qualification protocols (IQ, PQ and OQ).

Expertise
Our team of skilled technicians are experts at fixing analytical instruments. We have been working with Agilent, Waters, Perkin Elmer, Sciex, Shimadzu, and Thermo instruments for many years. We are constantly updating our knowledge with the latest technology and information to enable us to deliver the best support in the industry. We provide value added support and qualification/validation with all our instruments. We endeavor to develop strategic partnerships with our clients so that they consider our team as an extension of their employees. We are an affordable alternative to many normally expensive business projects.

Warranty and Support
We offer warranty on all refurbished instruments. Warranty starts at 3 months, with the option to extend the period. While your instrument is on warranty, we will provide Email (support@spectralabsci.com) and phone support and troubleshooting if you require assistance. If there is ever anything wrong with the equipment, we will replace it. Warranty begins from the date your equipment is successfully delivered. Any issues that occur during warranty should be reported immediately, describing the issues in as much detail as possible. Do not attempt to repair the equipment or involve a 3rd party without our express permission – this will result in voiding our warranty.

We will match our warranty duration with our competitors.

Installation and Training
If you purchase equipment from Spectralab Scientific, we offer installation and familiarization by our service technicians for an additional fee. If the instrument is particularly complicated, we highly recommend this service. If you’ve purchased less-complex equipment from us that you would like to install yourself, we can provide Team Viewer, email, or telephone support for no additional cost.

Large Inventory
We have a large inventory of equipment, components and parts in stock. You can find all the equipment you need for you lab in our store, as well as mix and match components for your systems.

International Shipping
We have sufficient export experience, including professional packing and proper documentation. Complementary shipping and handling can be applied upon immediate purchase. We include customs fees, brokerage, and duties for shipping to the US.

Price Matching & Discounts
Purchasing our refurbished instruments can save you anywhere from 30% to 90% compared to purchasing brand new. We will try to match or beat any price from our competitors, so feel free to inquire! Complementary shipping and handling can be applied upon immediate purchase.

Featured Products

  • New
    Shimadzu LCMS-8060 CL Triple Quadrupole Mass Spectrometer

    Shimadzu LCMS-8060 CL Triple Quadrupole Mass Spectrometer


    Condition:

    Refurbished

    Request QuoteView
  • New
    Agilent 8890 Gas Chromatograph G3540A

    Agilent 8890 Gas Chromatograph G3540A


    Condition:

    Refurbished

    Request QuoteView
  • NEW Thermo Dionex ASE 350 Accelerated Solvent Extractor


    Condition:

    Brand New

    Request QuoteView