Shimadzu Gas Chromatograph System High Performance and Expandability Merged at a higher level
High Performance
Improved design and innovative technology for all of our injectors, detectors and flow controllers equal or surpass our GC-2010 the high-end technology leader
Easy Operation
Large LCD, all digital gases control and auto-diagnostics inherited from the GC-
2010 – “The Most advanced, easy-to-use interface.
Flexibility
Use any column types for any analysis. Packed or capillary columns give you
the freedom to choose the best technique for your measurement. Fully integrated multiple valve systems are made simple for optimum performance for SystemGC custom GC products.
This Shimadzu GC-2014 GC with SFID and PID is in great condition. It has been tested at SpectraLab Scientific.
- Injection unit DINJ-W14
- DFID
- DTCD
- DAFC
- 7-port valve and 10-port valve
- Column oven coolant option
We have special expertise on analytical equipment. We encourage you to contact us to speak with one of our expert and professional team members.
Thanks for looking and we look forward to hearing from you!
Specifications
Column Oven |
|
Temperature range |
(Ambient + 10°C) ~ 400°C (using liquid CO2 gas*: -50°C ~400°C) |
Dimensions |
250 (W) × 360 (H) × 175 (D) mm |
Oven capacity |
15.8L |
Temperature accuracy |
Set value (K) ± 1% (calibration at 0.01°C increments) |
Temperature deviation |
2°C max. (on 200mm dia. circumference 30mm from rear) |
Temperature variation coefficient |
0.01°C/°C |
Temperature program steps |
Up to 20 (cooling program possible) |
Programmed rate setting range |
-250°C ~ 250°C/min |
Total time for all steps |
9999.99 minutes max |
Linear heating range |
30°C/min up to 150°C 20°C/min up to 250°C 10°C/min up to 380°C 7°C/min up to 400°C (at 25°C ambient temperature) |
Cooling rate |
300°C ~ 50°C in 6 min max. (at 25°C ambient temperature) |
Columns accepted |
Capillary columns: 2 Packed columns for GC14B: 4 (Glass columns: 2) |
Sample Injection Unit |
|
Temperature range |
Up to 400°C |
Heating settings |
1°C steps |
No. of units installed simultaneously |
Up to 3 units |
Sample injection unit types |
Dual packed, single packed, split/splitless, direct, direct (AMC) |
Carrier Gas Flow Controller |
|
For Packed / Dual |
|
Flow rate setting range |
0 ~ 100mL/min |
Programmable steps |
7 |
Programmed rate setting range |
400 ~ 400mL/min |
Correction function |
Maintains column flow rate during column oven heating |
For Capillary Split/Splitless, Direct |
|
(Split/splitless injection mode) |
|
Pressure setting range |
0 ~ 970kPa |
Programmable steps |
7 (pressure-decreasing program possible) |
Programmed rate setting range |
-400 ~ 400kPa/min |
Split ratio setting range |
0 ~ 9999.9 |
Total flow rate setting range |
0 ~ 1200mL/min |
Correction function |
Maintains column average linear velocity during column oven heating (for capillary only) |
(Pressure mode direct injection) |
|
Pressure setting range |
0 ~ 970kPa/min |
Programmable steps |
7 |
Programmed rate setting range |
-400 ~ 400kPa/min |
Correction function |
Maintains column average linear velocity during column oven heating |
(Flow-rate mode direct injection) |
|
Flow rate setting range |
0 ~ 1200mL/min |
Programmable steps |
7 |
Programmed rate setting range |
-400 ~ 400mL/min |
For Single Packed, Direct (AMC)
|
|
Flow rate setting range |
0~100mL/min |
Correction function
|
Maintenance column flow rate during column oven heating |
Detectors |
|
Temperature range |
400°C max. (FID, TCD) |
Temperature setting |
1°C steps |
No. of units installed simultaneously |
Up to 4 units (restricted depending on detector type) |
Flame Ionization Detector (FID) |
|
System |
Dual flow rate differential system |
Temperature range |
400°C max. |
Minimum detected quantity |
3pgC/s (dodecane) |
Dynamic range |
107 |
Nozzle
|
Quartz glass Standard: for packed, Option: for capillary |
Thermal Conductivity Detector (TCD) |
|
System |
Dual flow rate differential system |
Temperature range |
400°C max. |
Dynamic range |
105 |
Sensitivity |
40,000mV · mL/mg (built-in pre-amplifier, with 10 × application) |
System |
Fixed current system using 63Ni370MBq radiation source |
Temperature range |
400°C max. |
Minimum detected quantity |
0.1pg/s (γ-BHC) |
Dynamic range |
104 |
Flame Photometric Detector (FPD) |
|
Temperature range |
350°C max. |
Dynamic range |
P: 104 S: 103 |
Minimum detected quantity |
P: 0.5pgP/s (tributyl phosphate) S: 8pgS/s (dodecane thiol) |
Flame Thermionic Detector (FTD) |
(Two types, one for capillary and one for packed. The speci‑cation are the same.) |
Temperature range |
400°C max. |
Dynamic range |
N: 103 P: 103 |
Minimum detected quantity |
N: 0.4pgN/s (azobenzene) P: 0.05pgP/s (malathion) |