The Agilent 1290 Infinity II G7167A Multisampler can handle both vials and microtiter plates with ease and efficiency up to 1300 bar system pressure. This compact module has the capacity to house up to 6,144 samples, all inside the Agilent stack footprint – more than any single sampler from any other vendor – and the robotics to smoothly inject each into the chromatograph in turn. With Agilent’s dual-needle design, cycle time can be reduced efficiently using smart overlapped injections. And with our multi-wash capability, you can reduce carryover to less than 9 parts per million.
S/N: DEAC100265
Specifications:
Type | Specifications |
---|---|
Injection range | 0.1 – 20 µL in 0.1 µL increments (default) 0.1 – 40 µL in 0.1 µL increments if 40 µL loop is installed 0.1 – 120 µL in 0.1 µL increments with 1290 Infinity large volume injection kit (hardware modification required) pressure range up to 1200 bar 0.1 – 100 µL in 0.1 µL increments with 100 µL upgrade kit hardware (modification required) up to 1200 bar |
Precision | Typically <0.15 % RSD or SD<10 nL |
Pressure range | Up to 1200 bar > 500 µL Multidraw with 100 µL Analytical Head Up to 400 bar > 900 µL Analytical Head |
Sample viscosity range | 0.2 – 5 cp |
Sample capacity | 1H Drawer Up to 8 drawers and 16 positions Shallow well plates (MTP) 2H DrawerUp to 4 drawers and 8 positions MTP, deep well plates, vials, Eppendorf 3H DrawerUp to 2 drawers and 4 positions MTP, deep well plates, vials up to 6 mL, Eppendorf |
Injection cycle time | <10 s using following standard conditions: Default draw speed: 100 µL/min Default eject speed: 400 µL/min Injection volume: 1 µL |
Carry Over | <0.003 % (30 ppm) Multisampler Standard <0.0009 % (9 ppm) Multisampler Multiwash |
Control and data evaluation | Agilent Open Lab CDS Mass hunter Lab Advisor ICF for 3rd party SW control LC and CE Drivers |
Local Control | Agilent Instant Pilot (G4208A) |
Communications | Controller-area network (CAN),Local Area Network (LAN) ERI: ready, start, stop and shut-down signals |
Safety and maintenance | Extensive support for troubleshooting and maintenance is provided by the Instant Pilot, Agilent Lab Advisor, and the Chromatography Data System. Safety-related features are leak detection, safe leak handling, leak output signal for shutdown of pumping system, and low voltages in major maintenance areas. |