The ES2000 is a reach-in environmental chamber designed for high-demand testing, processing and conditioning.
-BT = Benchtop 12 cu.ft.
2 Shelves
Temperature range: 2 – 70°c (36-158°F)
Humidity Range: 10 – 96% RH
Max Shipping Weight: 261kg (575 LB)
Condensing Unit: 1/3 HP, R134a
Electrical Power: 208-230V / 1ph / 60hz
Voltage Utilization: 198 – 253
Full Load Amps: 21.3
Cord Set(Optional): NEMA 14-30
Total Heat of Rejection:
Normal Maximum
1876 W @ 25°c 1920 W @ 10°c
Ambient Temperature: 18°c (65°F) to 25°C (77°F)
Ambient Humidity: 45 to 55% RH
Operating Altitude: <2000m (6561 ft)
Operating Environment: Indoor
Temperature Uniformity: +/- 0.3°C (+/- .06°F)
Humidity Uniformity: +/- 3% RH
Interior Dimensions: 91 cm x 65 cm x 62 cm (W x D x H)
(36” x 25.5” x 24.5”)
Exterior Dimensions: 104 cm x 74 cm x 119 cm (W x D x H)
41” x 29” x 47”
Shelving: 2 Stainless Steel Shelves / 14 Available
Shelving Holding Capacity: 59 kg (130lb)
Shelf Area: 2.4 mˆ2 (6.5 ft ˆ2)
Interior Finish: 22 Gauge #4 304 Stainless Steel
Exterior and Finish: Steel with Grey Enamel Finish
Alarm Signals: Visual, Audible, Dry Contacts
Water Supply Requirements : < 0.5 ml/s (0.5 GPH) @ < 1.7 bar (25 psi) Water Quality Resistivity Limits: > 25 k ohms & < 15 M ohms
Water Supply/ Drain Connection: 0.25” OD Tube Quick Connect
Water Quality Equivalent Hardness: < 20 ppm or 1 GPG Water Quality pH level: > 6.7
Waste Water Drain 0.05” OD Stainless Steel
Typical Heat Up Times (30 sec. Door Open)
Ambient to + 37°C (99°F) < 5 Minutes
+37°C (99°F) to + 50°C (122°F) < 7 Minutes
+10°C (50°F) to + 50°C (122°F) < 15 Minutes
Typical Recovery to Setpoint (30 sec, door open)
@ + 50°C (122°F) < 2 Minutes
@ + 10°C (50°F) < 3 Minutes