Features
The ICS-5000 Reagent-Free™ IC (RFIC™) system with Eluent Generation is the world’s first capillary IC system. Developed with flexibility, modularity, and ease-of-use in mind, the ICS-5000 combines the highest sensitivity with convenience.
With the ability to analyze samples at capillary, microbore, or standard flow rates (or any combination of two, in a dual system), the system is simply Always Ready. The wide variety of ICS-5000 modules allows you to configure an ion chromatography system designed specifically to provide the solutions you need.
The Dionex ICS-5000 is the first system with an all-PEEK flow path capable of operating continuously at 5000 psi. Reagent-Free HPIC systems redefine the way ion chromatography is performed.
This Dionex Ics-5000 Ion Chromatography System includes:
- Dionex ICS-5000 Dual Pump (DP)
- Dionex ICS-5000 Eluent Generator (EG)
- Dionex ICS-5000 Detector/Chromatography Module
- Dionex AS-50 Autosampler
- A 90-day warranty
This refurbished Dionex Ics-5000 Ion Chromatography System is in great condition. It has been tested at SpectraLab Scientific.
We have special expertise on analytical equipment. We encourage you to contact us to speak with one of our expert and professional team members.
Thanks for looking and we look forward to hearing from you!
Dionex ICS-5000 Dual Pump(DP)
Specifications |
|
Type |
Dual-piston (in series), microprocessor-controlled, constant-stroke, variable-speed, patented isokinetic Eluent Precompression |
Construction |
Chemically inert, metal-free PEEK pump heads and flow path; compatible with aqueous eluents from pH 0–14 and reversed-phase solvents |
Pressure Range |
0–41 MPa (0–6000 psi) |
Flow Rate Range |
Analytical: 0.000–10.000 mL/min with settable flow increments at 0.001 mL/min Capillary: 0.001–3.000 mL/min with settable flow increments at 0.0001 mL/min |
Flow Rate Precision |
<0.1% |
Flow Rate Accuracy |
<0.1% at 14 MPa (2000 psi) |
Pressure Ripple |
Analytical: < 1.0% at 1.0 mL/min typical Capillary: < 0.2% (with damper) or < 1.0% (without damper) at 10 μL/min |
Pressure |
None required |
Vacuum Degasser |
Integrated degasser with 1 channel for isocratic pump or 4 channels for quaternary pump; optional |
Piston Seal Wash |
Automatic operation, standard |
Gradient Formation |
Quaternary low pressure (Analytical only) or electrolytic eluent generation at high pressure (Analytical and Capillary) |
RFIC-EG Gradients |
Analytical: 0.1–100 mM Capillary: 0.1–200 mM |
Gradient Profiles |
Any combination of an unlimited number of linear, convex, and concave positive and negative gradient profiles |
Gradient Proportioning Accuracy and Precision (Analytical only) |
±0.5% at 2 mL/min |
Gradient Mixing (Analytical only) |
Passive mixers for 2 mm and 4 mm ID columns, Optional |
Eluent Generation |
Eluent generation (RFIC-EG), optional |
Eluent On/Off Valve |
Electrically-actuated, standard |
Leak Sensor |
Optical, standard |
Dionex ICS-5000 Eluent Generator (EG)
Specification |
|
Minimum and Maximum Eluent Concentrations |
0.01–100 mM (Analytical) 0.01–200 mM (Capillary) |
Flow Rates |
0.100–3.000 mL/min (Analytical) 0.001–0.030 mL/min (Capillary) |
Eluent Types |
KOH, LiOH, NaOH |
Maximum Operating Pressure |
35 MPa (5000 psi) |
Maximum Solvent Concentration |
Cations: None |
Gradient Profiles |
Standard—any combination of an unlimited number of linear, convex, and concave positive and negative gradient profiles |
Number of Cartridges Supported |
Two—dual cartridge support |
Dionex ICS-5000 Detector/Chromatography Module
Specification |
|
DC models: |
Dual Temperature Model Only |
Dual Temperature Model: |
Upper Section Temp. Range: 10–40 °C |
Lower Zone: (Analytical) |
Injection valves: 1 or 2 ea., 6 or 10 port |
Cap IC Cube: (Optional – but required for capillary chemistries) |
1 or 2 IC Cube™ modules hold the following: |
Automation Manager: (Optional) |
Injection valves: Up to 2 ea., high pressure valves, either 6 or 10 port |
Leak detection: |
Optical leak sensor |
Dionex AS-50 Autosampler
Specification |
|
Injections Per Vial |
1 to 99 |
Vial Sizes Available |
0.3 mL polymer vials (P/N 055428, package of 100) 1.5 mL glass vials (P/N 055427, package of 100) 10 mL PolyVials (P/N 055058, package of 100) |
Variable Volume Range |
1 to 100 µL in 0.1 µL increments 100 to 1000 mL in 1 µL increments |
Injection Loop Size |
IC version: 25 µL standard; other sizes are available HPLC version: 20 µL standard; other sizes are available |
Injection |
Two-position, six-port, electrically-activated Rheodyne valve IC version: PEEK with Tefzel® rotor seal HPLC version: Stainless steel with Vespel® rotor seal |
Column Switching |
Two-position, ten-port, electrically-activated Rheodyne valve IC version: PEEK with Tefzel rotor seal HPLC version: Stainless steel with Vespel rotor seal |