Mettler Toledo XS105DU DualRange Analytical Balance with RS-P26 Printer.
Specifications:
Limit values | |
Maximum capacity | 120 g |
Readability | 0.1 mg |
Tare range (from…to) | 0 … 120 g |
Maximum capacity in fine range | 41 g |
Readability in fine range | 0.01 mg |
Repeatability (at nominal load) | 0.1 mg (100 g) |
Repeatability (at low load) | 0.05 mg (5 g) |
Repeatability in fine range (at low load) | 0.02 mg (5 g) |
Linearity deviation | 0.2 mg |
Eccentricity deviation (test load) 1) | 0.3 mg (50 g) |
Sensitivity offset (test weight) | 0.8 mg (100 g) |
Sensitivity temperature drift 2) | 0.00015%/°C |
Sensitivity stability 3) | 0.0002%/a |
Typical values | |
Repeatability (at low load) | 0.04 mg (5 g) |
Repeatability in fine range (at low load) | 0.01 mg (5 g) |
Linearity deviation | 0.06 mg |
Eccentricity deviation (test load) 1) | 0.075 mg (50 g) |
Sensitivity offset (test weight) | 0.2 mg (100 g) |
Minimum weight (according to USP) | 82 mg |
Minimum weight (according to USP) in fine range | 20 mg |
Minimum weight (U=1%, k=2) | 8.2 mg |
Minimum weight (U=1%, k=2) in fine range | 2 mg |
Settling time | 1.5 s |
Settling time in fine range | 3 s |
Interface update rate | 23 1/s |
Usable height of draft shield | 235 mm |
Weight of balance | 9.1 kg |
Number of built-in reference weights | 2 |
Dimensions | |
Balance dimensions (W × D × H) | 263 × 453 × 322 mm |
Weighing pan dimensions | 78 × 73 mm (W × D) |
Weights for routine testing | |
OIML CarePac
Weights |
100 g F2, 5 g E2
#11123002 |
ASTM CarePac
Weights |
100 g 1, 5 g 1
#11123102 |
1) According to OIML R76
2) In the temperature range 10 … 30 °C
3) After putting into operation for the first time, with the self-adjustment function activated (ProFACT or FACT)