Utilize the increased separation speed and resolution of the Thermo Scientific Vanquish UHPLC system to detect and quantify virtually any substance using the near-universal detection capabilities of Vanquish Charged Aerosol Detectors. With enhanced linear dynamic range and accurate, consistent response, this detector can measure low-level impurities and active pharmaceutical ingredients (APIs) simultaneously, as well as other compounds, regardless of chemical structure. One can measure virtually any nonvolatile and most semivolatile analytes with sub-nanogram sensitivity. The charged aerosol detector expands the capabilities for discovery and is a powerful addition to any chromatography lab.
Dimensions (L x W x H) | 192 mm x 420 mm x 620 mm (7.6 in. x 16.5 in. x 24.4 in.) Weight: 17.5 kg (38.6 lb.) |
Data Collection Rate | 100 Hz max |
Detection Mode | Charged Aerosol Detection |
Type | Charged Aerosol Detector |
Communications | 1 USB port (USB 2.0, “B” type) 1 USB hub with 3 ports (USB 2.0, “A” type) to connect further Vanquish modules |
GLP Features | Predictive Performance functions: Gas filter change interval, service monitoring period, system parameters logged in the Chromeleon Chromatography Data System Audit Trail. |
Analog Outputs | 0-1 V DC with full scale output range: 1 pA to 500 pA in 1-2-5 sequence (installable option) |
Gas Requirements | Nitrogen or Compressed Air (Nitrogen recommended) Inlet Pressure: 482-551 kPa (4.8-5.5 bar, 70-80 psi) |
Temperature Control | Evaporation temperature: Select: +35°C, +50°C, +70°C |
Inlet Gas Pressure | Electronically controlled pressure regulation system |
Filter | Filter Time Constant: 4 th order low-pass Bessel digital with time constraint based user selections: 0.1, 0.2, 0.5, 1.0, 2.0, 3.6, 5.0, 10.0 sec |
PC Connection | 1 USB port (USB 2.0, “B” type) 1 USB hub with 3 ports (USB 2.0, “A” type) to connect further Vanquish modules |
Range | Dynamic Range: Up to 4 orders of magnitude |
Safety Features | Power-up diagnostics, inlet gas over-pressure, high evaporation temp, pump flow shutoff, fluid leak detection and safe leak handling |
Description | Vanquish Charged Aerosol Detector F with Concentric Flow Nebulizer |
Electrical Requirements | 100-240 VAC, 50/60 Hz, max. 150 W/255 VA |
For Use With (Equipment) | Vanquish UHPLC System |
Unit Size | Each |